FDA Devices recall · Reported November 20, 2019
Visual-ICE utilizing the IceSeed 1.5 and IceRod 1.5 PLUS cryoablation needles could have issues with the needle clogging and/or inadequate freezing performance.
Galil Medical, Inc. recalled Visual-ICE Cryoablation System, FPRCH6000. The cryoablation system is a mobile console s… - a moderate-severity action.
Visual-ICE Cryoablation System, FPRCH6000. The cryoablation system is a mobile console s… was recalled by Galil Medical, Inc. in November 20, 2019. Reason: Visual-ICE utilizing the IceSeed 1.5 and IceRod 1.5 PLUS cryoablation needles could have issues with the need…. Check the official notice for the remedy. Verify recall #Z-0404-2020 with the FDA Devices before acting.
The recall
Galil Medical, Inc. issued this moderate-severity FDA Devices recall-Visual-ICE utilizing the IceSeed 1.5 and IceRod 1.5 PLUS cryoablation needles could have issues with the need….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0404-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0404-2020) was formally reported on November 20, 2019, with the manufacturer initiating the action on October 25, 2019. It is classified under Moderate severity (Class II), with a current status of Terminated. Galil Medical, Inc. is listed as the recalling firm, operating out of Saint Paul, MN. Federal records list the affected scope as 49 systems.
The documented reason for this recall is: Visual-ICE utilizing the IceSeed 1.5 and IceRod 1.5 PLUS cryoablation needles could have issues with the needle clogging and/or inadequate freezing performance. Distribution data in the federal record shows the product reached: AK, AL, AR, AZ, CA, FL, GA, HI, KY, MA, MI, MN, NC, ND, NY, OH, PA, SC, TN, UT, VA, Puerto Rico OUS to include: Argentina, Canada, Taiwan, United Kingdom. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
49 systems
Related Recalls
6
3 from same agency
Visual-ICE Cryoablation System, FPRCH6000. The cryoablation system is a mobile console system intended for cryoablative tissue destruction using a minimally invasive procedure. The system is computer-controlled with a touch screen user interface that allows the user to control and monitor the procedure. The therapy delivered by the system is based on the Joule-Thomson effect displayed by compressed gases. The Visual-ICE System uses high-pressure argon gas that circulates through closed-tip cryoablation needles to induce tissue freezing. Active tissue thawing is achieved by circulating helium gas through the needles or, alternatively , by the use of Galil Medical I-Thaw technology in which a heating element inside the cryoablation needle can be energized to cause thawing.
Visual-ICE utilizing the IceSeed 1.5 and IceRod 1.5 PLUS cryoablation needles could have issues with the needle clogging and/or inadequate freezing performance.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0404-2020 |
| Date reported | November 20, 2019 |
| Date initiated | October 25, 2019 |
| Recalling firm | Galil Medical, Inc. |
| Firm location | Saint Paul, MN |
| Affected scope | 49 systems |
| Distribution | AK, AL, AR, AZ, CA, FL, GA, HI, KY, MA, MI, MN, NC, ND, NY, OH, PA, SC, TN, UT, VA, Puerto Rico OUS to include: Argentina, Canada, Taiwan, United Kingdom |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 20, 2019. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.