Severity
Moderate
FDA Devices recall · Reported December 18, 2013
As part of routine manufacturing process monitoring by Zimmer, it was discovered that the Tibial Base and Talar Components from the Total Ankle system have the potential of contai…
Zimmer, Inc. recalled Zimmer Trabecular Metal Total Ankle Talar Component Right Sizes 1-6 Total ankle arthro… - a moderate-severity action.
Zimmer Trabecular Metal Total Ankle Talar Component Right Sizes 1-6 Total ankle arthro… was recalled by Zimmer, Inc. in December 18, 2013. Reason: As part of routine manufacturing process monitoring by Zimmer, it was discovered that the Tibial Base and Tal…. Check the official notice for the remedy. Verify recall #Z-0476-2014 with the FDA Devices before acting.
The recall
Zimmer, Inc. issued this moderate-severity FDA Devices recall-As part of routine manufacturing process monitoring by Zimmer, it was discovered that the Tibial Base and Tal….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0476-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0476-2014) was formally reported on December 18, 2013, with the manufacturer initiating the action on November 12, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Zimmer, Inc. is listed as the recalling firm, operating out of Warsaw, IN. Federal records list the affected scope as 265 units (US) 114 units (outside US).
The documented reason for this recall is: As part of routine manufacturing process monitoring by Zimmer, it was discovered that the Tibial Base and Talar Components from the Total Ankle system have the potential of containing elevated levels of manufacturing ma… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) and Internationally to Australia, Canada, China, Finland, Germany, and Italy.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
265 units (US) 114 units (outside US)
Related Recalls
6
3 from same agency
Zimmer Trabecular Metal Total Ankle Talar Component Right Sizes 1-6 Total ankle arthroplasty is intended to provide a patient with limited mobility by restoring alignment, reducing pain and preserving the flexion/extension movement within the ankle joint.The Zimmer Trabecular Metal Total Ankle is indicated as a total ankle replacement in primary or revision surgery for patients with: " Rheumatoid arthritis. " Post-traumatic arthritis. " Degenerative arthritis. This device is intended for cemented use only.
As part of routine manufacturing process monitoring by Zimmer, it was discovered that the Tibial Base and Talar Components from the Total Ankle system have the potential of containing elevated levels of manufacturing materials residue on the implants. The specific material was found to be mass media utilized during the manufacturing process to provide the uniform surface finish. This mass media
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0476-2014 |
| Date reported | December 18, 2013 |
| Date initiated | November 12, 2013 |
| Recalling firm | Zimmer, Inc. |
| Firm location | Warsaw, IN |
| Affected scope | 265 units (US) 114 units (outside US) |
| Distribution | Worldwide Distribution - USA (nationwide) and Internationally to Australia, Canada, China, Finland, Germany, and Italy. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 18, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.