Severity
Moderate
FDA Devices recall · Reported February 28, 2018
FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and …
Fujifilm Medical Systems U.S.A., Inc. recalled Aspire HD (for FDR-1000AWS) The Fujifilm Digital Mammography System, Aspire HD (FDR MS… - a moderate-severity action.
Aspire HD (for FDR-1000AWS) The Fujifilm Digital Mammography System, Aspire HD (FDR MS… was recalled by Fujifilm Medical Systems U.S.A., Inc. in February 28, 2018. Reason: FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The…. Check the official notice for the remedy. Verify recall #Z-0659-2018 with the FDA Devices before acting.
The recall
Fujifilm Medical Systems U.S.A., Inc. issued this moderate-severity FDA Devices recall-FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0659-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0659-2018) was formally reported on February 28, 2018, with the manufacturer initiating the action on August 4, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Fujifilm Medical Systems U.S.A., Inc. is listed as the recalling firm, operating out of Stamford, CT. Federal records list the affected scope as 71 units (68 units Domestic. 3 units Foreign) in total.
The documented reason for this recall is: FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number t… Distribution data in the federal record shows the product reached: Worldwide Distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,861 medical devices recalls on record
Of 88,001 recalls with an agency hazard classification, 19,020 are high severity, 64,468 moderate, and 4,513 low. A further 16,157 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
71 units (68 units Domestic. 3 units Foreign) in total
Related Recalls
6
6 from same agency
Aspire HD (for FDR-1000AWS) The Fujifilm Digital Mammography System, Aspire HD (FDR MS-l000), generates full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.
FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0659-2018 |
| Date reported | February 28, 2018 |
| Date initiated | August 4, 2017 |
| Recalling firm | Fujifilm Medical Systems U.S.A., Inc. |
| Firm location | Stamford, CT |
| Affected scope | 71 units (68 units Domestic. 3 units Foreign) in total |
| Distribution | Worldwide Distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Two federal-filing-derived peer sets for this notice, both outside the same recalling firm and filing state so the neighborhoods are not firm or geographic containment.
Same archive severity grade (Moderate), closest by report date among other firms and states.
Same product category, closest by report date among other firms and states (deduped against the severity set).
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
Showing 5 of 40909 other recalls in this category.
Boston Scientific Corporation · 2026-07-29
Boston Scientific Corporation · 2026-07-29
AVID Medical, Inc. · 2026-07-29
Penumbra, Inc. · 2026-07-29
Alphatec Spine, Inc. · 2026-07-29
Compare this recall with FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 28, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of August 2026.