Severity
Moderate
FDA Devices recall · Reported March 7, 2018
It the WiFi connection between the SkyPlate detector and HP transfer point is weak, an image may fail to transfer from the SkyPlate detector to the system. The image remains in t…
Philips Electronics North America Corporation recalled digital x-ray detector ProGrade R1 - solid state X ray imager (flat panel/digital imag… - a moderate-severity action.
digital x-ray detector ProGrade R1 - solid state X ray imager (flat panel/digital imag… was recalled by Philips Electronics North America Corporation in March 7, 2018. Reason: It the WiFi connection between the SkyPlate detector and HP transfer point is weak, an image may fail to tran…. Check the official notice for the remedy. Verify recall #Z-0704-2018 with the FDA Devices before acting.
The recall
Philips Electronics North America Corporation issued this moderate-severity FDA Devices recall-It the WiFi connection between the SkyPlate detector and HP transfer point is weak, an image may fail to tran….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0704-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0704-2018) was formally reported on March 7, 2018, with the manufacturer initiating the action on December 11, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips Electronics North America Corporation is listed as the recalling firm, operating out of Andover, MA. Federal records list the affected scope as 157.
The documented reason for this recall is: It the WiFi connection between the SkyPlate detector and HP transfer point is weak, an image may fail to transfer from the SkyPlate detector to the system. The image remains in the memory of the detector, but cannot be… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) Distribution, Internationally to Canada.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,861 medical devices recalls on record
Of 88,001 recalls with an agency hazard classification, 19,020 are high severity, 64,468 moderate, and 4,513 low. A further 16,157 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
157
Related Recalls
6
6 from same agency
digital x-ray detector ProGrade R1 - solid state X ray imager (flat panel/digital imager) As a part of a radiographic system, the Philips ProGrade is intended to acquire, process, store, display, and export digital radiographic images. The Philips ProGrade is suitable for all routine radiographic examinations, including specialist area like intensive care, trauma, or pediatric work, excluding mammography. United States only: The Eleva Workspot is not intended for fluoroscopy and angiography.
It the WiFi connection between the SkyPlate detector and HP transfer point is weak, an image may fail to transfer from the SkyPlate detector to the system. The image remains in the memory of the detector, but cannot be transferred wirelessly or by use of the backup cable. To continue, the operator can reset the SkyPlate detector by removing its batteries, but the acquired image is lost and a re-take is necessary.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0704-2018 |
| Date reported | March 7, 2018 |
| Date initiated | December 11, 2017 |
| Recalling firm | Philips Electronics North America Corporation |
| Firm location | Andover, MA |
| Affected scope | 157 |
| Distribution | Worldwide Distribution - USA (nationwide) Distribution, Internationally to Canada. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Two federal-filing-derived peer sets for this notice, both outside the same recalling firm and filing state so the neighborhoods are not firm or geographic containment.
Same archive severity grade (Moderate), closest by report date among other firms and states.
Same product category, closest by report date among other firms and states (deduped against the severity set).
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
Showing 5 of 40909 other recalls in this category.
Boston Scientific Corporation · 2026-07-29
Boston Scientific Corporation · 2026-07-29
AVID Medical, Inc. · 2026-07-29
Penumbra, Inc. · 2026-07-29
Alphatec Spine, Inc. · 2026-07-29
Compare this recall with FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported March 7, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of August 2026.