PlainRecalls
CriticalClass IOngoing

FDA Devices recall · Reported February 7, 2024

EvAir CPRSR Kit Upgrade: a) DISS/CEE 7/7 230V 50/60 HZ, Model Number M1230849; b) DISS/NEMA 5-15 115V 60 HZ, Model Number M1230847; c) NIST/BS 1363 230V 50/60 HZ, Model Number M1230849; d) NIST/BS 546 230V 50/60 HZ, Model Number M1230849; e) NIST/CEE 7/7 230V 50/60 HZ, Model Number M1230849;

GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestati…

Recall #
Z-0776-2024
Affected scope
1294 units
Initiated
December 29, 2023
Verify with FDA Devices →
View my saved recalls

Datex-Ohmeda, Inc. recalled EvAir CPRSR Kit Upgrade: a) DISS/CEE 7/7 230V 50/60 HZ, Model Number M1230849; b) DI… - a critical-severity action.

EvAir CPRSR Kit Upgrade: a) DISS/CEE 7/7 230V 50/60 HZ, Model Number M1230849; b) DI… was recalled by Datex-Ohmeda, Inc. in February 7, 2024. Reason: GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) option…. Check the official notice for the remedy. Verify recall #Z-0776-2024 with the FDA Devices before acting.

The recall

Datex-Ohmeda, Inc. issued this critical-severity FDA Devices recall-GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) option….

Critical
severity level
1K units
affected scope
Class I
classification
February 7, 2024
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0776-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified high severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

EvAir CPRSR Kit Upgrade: a) DISS/CEE 7/7 230V 50/60 HZ, Model Number M1230849; b) DISS/NEMA 5-15 115V 60 HZ, Model Number M1230847; c) NIST/BS 1363 230V 50/60 HZ, Model Number M1230849; d) NIST/BS 546 230V 50/60 HZ, Model Number M1230849; e) NIST/CEE 7/7 230V 50/60 HZ, Model Number M1230849;

Reason for recall

GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions,GE HealthCare is taking this action to further reduce the potential for patient exposure.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Critical (Class I)
Status Ongoing
Recall number Z-0776-2024
Date reported February 7, 2024
Date initiated December 29, 2023
Recalling firm Datex-Ohmeda, Inc.
Firm location Madison, WI
Affected scope 1294 units
Distribution Worldwide

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

1294 units units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0776-2024) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • This is a Class I (high-risk) recall, stop using the product immediately and follow the disposal or return instructions. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
EvAir CPRSR Kit Upgrade: a) DISS/CEE 7/7 230V 50/60 HZ, Model Number M1230849; b) DISS/NEMA 5-15 115V 60 HZ, Model Number M1230847; c) NIST/BS 1363 230V 50/60 HZ, Model Number M1230849; d) NIST/BS 546 230V 50/60 HZ, Model Number M1230849; e) NIST/CEE 7/7 230V 50/60 HZ, Model Number M1230849;. Recalled by Datex-Ohmeda, Inc.. Units affected: 1294 units.
Why was this product recalled?
GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions,GE HealthCare is taking this action to further reduce the potential for patient exposure.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 7, 2024. Severity: Critical. Recall number: Z-0776-2024.
Where was the recalled product distributed?
Distribution: Worldwide.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported February 7, 2024. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026