Custom procedural convenience kits and trays, cardiac, labeled as: a) CENTRAL LINE INSERTION PACK, kit number LLCE52; b) Adult Central Line Insertion Supply Kit, kit number UICL44N; c) Pediatric Central Line Insertion, kit number UIPL77K

Reported: February 7, 2024 Initiated: November 29, 2023 #Z-0819-2024

Product Description

Custom procedural convenience kits and trays, cardiac, labeled as: a) CENTRAL LINE INSERTION PACK, kit number LLCE52; b) Adult Central Line Insertion Supply Kit, kit number UICL44N; c) Pediatric Central Line Insertion, kit number UIPL77K

Reason for Recall

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Details

Units Affected
290 units
Distribution
US Nationwide distribution.
Location
Temple Terrace, FL

Frequently Asked Questions

What product was recalled?
Custom procedural convenience kits and trays, cardiac, labeled as: a) CENTRAL LINE INSERTION PACK, kit number LLCE52; b) Adult Central Line Insertion Supply Kit, kit number UICL44N; c) Pediatric Central Line Insertion, kit number UIPL77K. Recalled by American Contract Systems, Inc.. Units affected: 290 units.
Why was this product recalled?
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 7, 2024. Severity: Moderate. Recall number: Z-0819-2024.