Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; b) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; c) MINI MINOR AGH ASC PACK, kit number AGMM60M; d) OUTPATIENT PORT ACCESS KIT 307907, kit number AHPO55J; e) General Pack, kit number BBGP31C; f) FRACTURE TABLE PACK, kit number FDFT75N; g) Arthroscopy Pack, kit number HCAR04K; h) BARIATRIC PACK, kit number HGBA92F; i) LAPAROTOMY PACK, ki

Reported: February 7, 2024 Initiated: November 29, 2023 #Z-0825-2024 7881 units units

American Contract Systems, Inc. issued this FDA Devices recall on February 7, 2024. Classified as Moderate severity (Class II). Approximately 7881 units units are affected. The recall was issued because: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected …. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-0825-2024) was formally reported on February 7, 2024, with the manufacturer initiating the action on November 29, 2023. It is classified under Moderate severity (Class II), with a current status of Ongoing. American Contract Systems, Inc. is listed as the recalling firm, operating out of Temple Terrace, FL. Federal records indicate 7881 units units are affected.

The documented reason for this recall is: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" com… Distribution data in the federal record shows the product reached: US Nationwide distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 2 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

7881 units

Related Recalls

6

6 from same agency

Product Description

Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; b) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; c) MINI MINOR AGH ASC PACK, kit number AGMM60M; d) OUTPATIENT PORT ACCESS KIT 307907, kit number AHPO55J; e) General Pack, kit number BBGP31C; f) FRACTURE TABLE PACK, kit number FDFT75N; g) Arthroscopy Pack, kit number HCAR04K; h) BARIATRIC PACK, kit number HGBA92F; i) LAPAROTOMY PACK, kit number HNLP12J; j) MIDLINE CHANGE KIT, kit number HSMD20E; k) Minor General, kit number JSMI10I; l) ENDO VASCULAR PACK, kit number LVEV61; m) MAJOR PACK, kit number LVMJ21; n) HEAD AND NECK PACK, kit number MMHN37F; o) KIT, MINOR U BAR, kit number MMNU34E; p) KIT, MAJOR UNIVERSAL, kit number MMUN28E; q) VAG PLASTY PACK, kit number OSVP16; r) DR L MAJOR, kit number SALM40H; s) DR L MAJOR, kit number SALM40I; t) MINOR BASIN SAH, kit number SAMB25E; u) MAJOR SAH, kit number SAMJ36G; v) MINOR PACK, kit number SMMN65; w) General Pack, kit number THGN59D; x) MINOR PACK, kit number TPMN15; y) MAJOR UNIVERSAL PACK, kit number TPUN48

Reason for Recall

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Details

Units Affected
7881 units
Distribution
US Nationwide distribution.
Location
Temple Terrace, FL

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-0825-2024
Date reported February 7, 2024
Date initiated November 29, 2023
Recalling firm American Contract Systems, Inc.
Units affected 7881 units
Distribution US Nationwide distribution.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

7881 units units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; b) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; c) MINI MINOR AGH ASC PACK, kit number AGMM60M; d) OUTPATIENT PORT ACCESS KIT 307907, kit number AHPO55J; e) General Pack, kit number BBGP31C; f) FRACTURE TABLE PACK, kit number FDFT75N; g) Arthroscopy Pack, kit number HCAR04K; h) BARIATRIC PACK, kit number HGBA92F; i) LAPAROTOMY PACK, kit number HNLP12J; j) MIDLINE CHANGE KIT, kit number HSMD20E; k) Minor General, kit number JSMI10I; l) ENDO VASCULAR PACK, kit number LVEV61; m) MAJOR PACK, kit number LVMJ21; n) HEAD AND NECK PACK, kit number MMHN37F; o) KIT, MINOR U BAR, kit number MMNU34E; p) KIT, MAJOR UNIVERSAL, kit number MMUN28E; q) VAG PLASTY PACK, kit number OSVP16; r) DR L MAJOR, kit number SALM40H; s) DR L MAJOR, kit number SALM40I; t) MINOR BASIN SAH, kit number SAMB25E; u) MAJOR SAH, kit number SAMJ36G; v) MINOR PACK, kit number SMMN65; w) General Pack, kit number THGN59D; x) MINOR PACK, kit number TPMN15; y) MAJOR UNIVERSAL PACK, kit number TPUN48. Recalled by American Contract Systems, Inc.. Units affected: 7881 units.
Why was this product recalled?
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 7, 2024. Severity: Moderate. Recall number: Z-0825-2024.
Where was the recalled product distributed?
Distribution: US Nationwide distribution..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0825-2024) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).