Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR NEURO PACK, kit number AGNE20J; b) ADULT CRANIOTOMY PACK, kit number LLCR99; c) LAMINECTOMY PACK, kit number LLLC26; d) KIT, CRANIOTOMY, kit number MMCR11J; e) KIT, NEURO SPINE, kit number MMNS20H; f) IR NEURO PACK, kit number OSIN77; g) NEURO SAH, kit number SANE32H; h) NEURO SAH, kit number SANE32I; i) S78FCH CRANI PACK (PS 132432), kit number UICR10U; j) CRANI PACK (PS039498), kit number
Product Description
Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR NEURO PACK, kit number AGNE20J; b) ADULT CRANIOTOMY PACK, kit number LLCR99; c) LAMINECTOMY PACK, kit number LLLC26; d) KIT, CRANIOTOMY, kit number MMCR11J; e) KIT, NEURO SPINE, kit number MMNS20H; f) IR NEURO PACK, kit number OSIN77; g) NEURO SAH, kit number SANE32H; h) NEURO SAH, kit number SANE32I; i) S78FCH CRANI PACK (PS 132432), kit number UICR10U; j) CRANI PACK (PS039498), kit number UICR33AW; k) Laminectomy, kit number UILM77AR; l) DIAGNOSTIC NEURO IR TRAY, kit number WENE21D; m) DIAGNOSTIC NEURO IR TRAY, kit number WENE21F; n) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H; o) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H-01; p) NEURO IR TRAY, kit number WENI56F; q) NEURO IR TRAY, kit number WENI56H; r) NEURO IR TRAY, kit number WENI56H-01
Reason for Recall
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Details
- Recalling Firm
- American Contract Systems, Inc.
- Units Affected
- 2717 units
- Distribution
- US Nationwide distribution.
- Location
- Temple Terrace, FL
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11