Severity
Moderate
FDA Devices recall · Reported February 7, 2024
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting doc…
American Contract Systems, Inc. recalled Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR … - a moderate-severity action.
Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR … was recalled by American Contract Systems, Inc. in February 7, 2024. Reason: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and …. Check the official notice for the remedy. Verify recall #Z-0827-2024 with the FDA Devices before acting.
The recall
American Contract Systems, Inc. issued this moderate-severity FDA Devices recall-During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0827-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0827-2024) was formally reported on February 7, 2024, with the manufacturer initiating the action on November 29, 2023. It is classified under Moderate severity (Class II), with a current status of Ongoing. American Contract Systems, Inc. is listed as the recalling firm, operating out of Temple Terrace, FL. Federal records list the affected scope as 2717 units.
The documented reason for this recall is: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" com… Distribution data in the federal record shows the product reached: US Nationwide distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
2717 units
Related Recalls
6
3 from same agency
Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR NEURO PACK, kit number AGNE20J; b) ADULT CRANIOTOMY PACK, kit number LLCR99; c) LAMINECTOMY PACK, kit number LLLC26; d) KIT, CRANIOTOMY, kit number MMCR11J; e) KIT, NEURO SPINE, kit number MMNS20H; f) IR NEURO PACK, kit number OSIN77; g) NEURO SAH, kit number SANE32H; h) NEURO SAH, kit number SANE32I; i) S78FCH CRANI PACK (PS 132432), kit number UICR10U; j) CRANI PACK (PS039498), kit number UICR33AW; k) Laminectomy, kit number UILM77AR; l) DIAGNOSTIC NEURO IR TRAY, kit number WENE21D; m) DIAGNOSTIC NEURO IR TRAY, kit number WENE21F; n) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H; o) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H-01; p) NEURO IR TRAY, kit number WENI56F; q) NEURO IR TRAY, kit number WENI56H; r) NEURO IR TRAY, kit number WENI56H-01
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0827-2024 |
| Date reported | February 7, 2024 |
| Date initiated | November 29, 2023 |
| Recalling firm | American Contract Systems, Inc. |
| Firm location | Temple Terrace, FL |
| Affected scope | 2717 units |
| Distribution | US Nationwide distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 7, 2024. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.