Severity
Moderate
FDA Devices recall · Reported February 7, 2024
Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could …
Covidien LP recalled Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 S… - a moderate-severity action.
Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 S… was recalled by Covidien LP in February 7, 2024. Reason: Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma l…. Check the official notice for the remedy. Verify recall #Z-0844-2024 with the FDA Devices before acting.
The recall
Covidien LP issued this moderate-severity FDA Devices recall-Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma l….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0844-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0844-2024) was formally reported on February 7, 2024, with the manufacturer initiating the action on December 21, 2023. It is classified under Moderate severity (Class II), with a current status of Ongoing. Covidien LP is listed as the recalling firm, operating out of Mansfield, MA. Federal records list the affected scope as 46,812 US; 68,976 OUS.
The documented reason for this recall is: Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence… Distribution data in the federal record shows the product reached: Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries of Australia, Bahrai…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
46,812 US; 68,976 OUS
Related Recalls
6
3 from same agency
Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT 88861919-71 SURGILON* 1 BLK 7X75CM PCT
Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0844-2024 |
| Date reported | February 7, 2024 |
| Date initiated | December 21, 2023 |
| Recalling firm | Covidien LP |
| Firm location | Mansfield, MA |
| Affected scope | 46,812 US; 68,976 OUS |
| Distribution | Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries of Australia, Bahrain, Belgium, Bulgaria… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 7, 2024. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.