Severity
Moderate
FDA Devices recall · Reported April 10, 2013
Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static elec…
Medtronic Neuromodulation recalled Medtronic Prostiva RF Therapy Generator. The Prostiva RF system is indicated for the t… - a moderate-severity action.
Medtronic Prostiva RF Therapy Generator. The Prostiva RF system is indicated for the t… was recalled by Medtronic Neuromodulation in April 10, 2013. Reason: Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930…. Check the official notice for the remedy. Verify recall #Z-1031-2013 with the FDA Devices before acting.
The recall
Medtronic Neuromodulation issued this moderate-severity FDA Devices recall-Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1031-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1031-2013) was formally reported on April 10, 2013, with the manufacturer initiating the action on February 27, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Medtronic Neuromodulation is listed as the recalling firm, operating out of Minneapolis, MN. Federal records list the affected scope as 848 (696 US, 152 OUS).
The documented reason for this recall is: Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static electricity) to the footswitch connector por… Distribution data in the federal record shows the product reached: Nationwide Distribution including Puerto Rico and all States in continental USA except ID, MT, and NH.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
848 (696 US, 152 OUS)
Related Recalls
6
3 from same agency
Medtronic Prostiva RF Therapy Generator. The Prostiva RF system is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostate sizes between 20 and 50 cm3.
Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static electricity) to the footswitch connector port may toggle the RF output on or off. If this were to occur, it could result in unintentionally starting or stopping the lesion process, depending on the current status of the equi
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1031-2013 |
| Date reported | April 10, 2013 |
| Date initiated | February 27, 2013 |
| Recalling firm | Medtronic Neuromodulation |
| Firm location | Minneapolis, MN |
| Affected scope | 848 (696 US, 152 OUS) |
| Distribution | Nationwide Distribution including Puerto Rico and all States in continental USA except ID, MT, and NH. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 10, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.