Severity
Moderate
FDA Devices recall · Reported April 10, 2013
Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris…
Beckman Coulter Inc. recalled Access 2 Immunoassay System; P/N 107405S 2/12 The Access Immunoassay Systems are in vi… - a moderate-severity action.
Access 2 Immunoassay System; P/N 107405S 2/12 The Access Immunoassay Systems are in vi… was recalled by Beckman Coulter Inc. in April 10, 2013. Reason: Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that …. Check the official notice for the remedy. Verify recall #Z-1033-2013 with the FDA Devices before acting.
The recall
Beckman Coulter Inc. issued this moderate-severity FDA Devices recall-Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1033-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1033-2013) was formally reported on April 10, 2013, with the manufacturer initiating the action on February 12, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Beckman Coulter Inc. is listed as the recalling firm, operating out of Brea, CA. Federal records list the affected scope as 3500 units.
The documented reason for this recall is: Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator… Distribution data in the federal record shows the product reached: Worldwide distribution: USA (nationwide) and countries of: Aurstria, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Djibouti, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, N…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
3500 units
Related Recalls
6
3 from same agency
Access 2 Immunoassay System; P/N 107405S 2/12 The Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.
Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley rubbing against the light shield can fall into the reaction vessels (RVs) and present a risk of erroneous test results.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1033-2013 |
| Date reported | April 10, 2013 |
| Date initiated | February 12, 2013 |
| Recalling firm | Beckman Coulter Inc. |
| Firm location | Brea, CA |
| Affected scope | 3500 units |
| Distribution | Worldwide distribution: USA (nationwide) and countries of: Aurstria, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Djibouti, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Netherlands, Poland, … |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 10, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.