Severity
Moderate
FDA Devices recall · Reported January 21, 2026
Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the…
C-rad Positioning Ab recalled Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-002… - a moderate-severity action.
Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-002… was recalled by C-rad Positioning Ab in January 21, 2026. Reason: Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch …. Check the official notice for the remedy. Verify recall #Z-1092-2026 with the FDA Devices before acting.
The recall
C-rad Positioning Ab issued this moderate-severity FDA Devices recall-Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1092-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1092-2026) was formally reported on January 21, 2026, with the manufacturer initiating the action on October 9, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. C-rad Positioning Ab is listed as the recalling firm, operating out of Uppsala. Federal records list the affected scope as 189.
The documented reason for this recall is: Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the … Distribution data in the federal record shows the product reached: The software module cSRS has been installed in 37 countries Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, CA, CT, FL, MO, NC, NM, NY, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argenti…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
189
Related Recalls
6
3 from same agency
Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable
Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1092-2026 |
| Date reported | January 21, 2026 |
| Date initiated | October 9, 2025 |
| Recalling firm | C-rad Positioning Ab |
| Firm location | Uppsala |
| Affected scope | 189 |
| Distribution | The software module cSRS has been installed in 37 countries Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, CA, CT, FL, MO, NC, NM, NY, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Australia, Austr… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported January 21, 2026. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.