Severity
Moderate
FDA Devices recall · Reported January 28, 2026
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no cont…
Raysearch Laboratories Ab recalled Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 10.1.1 S… - a moderate-severity action.
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 10.1.1 S… was recalled by Raysearch Laboratories Ab in January 28, 2026. Reason: Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of…. Check the official notice for the remedy. Verify recall #Z-1107-2026 with the FDA Devices before acting.
The recall
Raysearch Laboratories Ab issued this moderate-severity FDA Devices recall-Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1107-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1107-2026) was formally reported on January 28, 2026, with the manufacturer initiating the action on November 28, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Raysearch Laboratories Ab is listed as the recalling firm, operating out of Stockholm. Federal records list the affected scope as 0.
The documented reason for this recall is: Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override… Distribution data in the federal record shows the product reached: Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
0
Related Recalls
6
3 from same agency
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 10.1.1 Software Version: RayStation 10B SP1 Product Description: Radiation Therapy Treatment Planning System Component: No
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1107-2026 |
| Date reported | January 28, 2026 |
| Date initiated | November 28, 2025 |
| Recalling firm | Raysearch Laboratories Ab |
| Firm location | Stockholm |
| Affected scope | 0 |
| Distribution | Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported January 28, 2026. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.