Severity
Moderate
FDA Devices recall · Reported March 11, 2020
Some lots ending in "17", "18", "19", or higher may have cuvette defects allowing water from the water bath to contaminate the interior of the cuvette, which may falsely elevate o…
Siemens Healthcare Diagnostics, Inc. recalled Atellica CH Reaction Ring Segment, SMN 11099326, UDI 00630414596143 - a moderate-severity action.
Atellica CH Reaction Ring Segment, SMN 11099326, UDI 00630414596143 was recalled by Siemens Healthcare Diagnostics, Inc. in March 11, 2020. Reason: Some lots ending in "17", "18", "19", or higher may have cuvette defects allowing water from the water bath t…. Check the official notice for the remedy. Verify recall #Z-1421-2020 with the FDA Devices before acting.
The recall
Siemens Healthcare Diagnostics, Inc. issued this moderate-severity FDA Devices recall-Some lots ending in "17", "18", "19", or higher may have cuvette defects allowing water from the water bath t….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1421-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1421-2020) was formally reported on March 11, 2020, with the manufacturer initiating the action on January 12, 2020. It is classified under Moderate severity (Class II), with a current status of Ongoing. Siemens Healthcare Diagnostics, Inc. is listed as the recalling firm, operating out of Tarrytown, NY. Federal records list the affected scope as 3559.
The documented reason for this recall is: Some lots ending in "17", "18", "19", or higher may have cuvette defects allowing water from the water bath to contaminate the interior of the cuvette, which may falsely elevate or depress sample results. Distribution data in the federal record shows the product reached: Domestic distribution nationwide. Worldwide foreign distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
3559
Related Recalls
6
3 from same agency
Atellica CH Reaction Ring Segment, SMN 11099326, UDI 00630414596143
Some lots ending in "17", "18", "19", or higher may have cuvette defects allowing water from the water bath to contaminate the interior of the cuvette, which may falsely elevate or depress sample results.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1421-2020 |
| Date reported | March 11, 2020 |
| Date initiated | January 12, 2020 |
| Recalling firm | Siemens Healthcare Diagnostics, Inc. |
| Firm location | Tarrytown, NY |
| Affected scope | 3559 |
| Distribution | Domestic distribution nationwide. Worldwide foreign distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported March 11, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.