t:slim X2 Insulin Pump with Interoperable Technology
Product Description
t:slim X2 Insulin Pump with Interoperable Technology
Reason for Recall
A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.
Details
- Recalling Firm
- Tandem Diabetes Care, Inc.
- Units Affected
- 63 units
- Distribution
- US: AZ, CA, CO, CT, FL, HI, IA, ID, IL, IN, MD, ME, MI, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, PA, TX, UT, VA, WA, WI OUS: N/A
- Location
- San Diego, CA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11