Severity
Moderate
FDA Devices recall · Reported May 26, 2021
the firm discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over time, the shorter pin may not engage in either the upper or lower body,…
WillowWood Global Llc recalled Alpha Interlock Suspension System Lanyard Body: An external prosthetic device. The Alpha … - a moderate-severity action.
Alpha Interlock Suspension System Lanyard Body: An external prosthetic device. The Alpha … was recalled by WillowWood Global Llc in May 26, 2021. Reason: the firm discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over t…. Check the official notice for the remedy. Verify recall #Z-1594-2021 with the FDA Devices before acting.
The recall
WillowWood Global Llc issued this moderate-severity FDA Devices recall-the firm discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over t….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1594-2021 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1594-2021) was formally reported on May 26, 2021, with the manufacturer initiating the action on April 20, 2021. It is classified under Moderate severity (Class II), with a current status of Ongoing. WillowWood Global Llc is listed as the recalling firm, operating out of Mount Sterling, OH. Federal records list the affected scope as 815 parts.
The documented reason for this recall is: the firm discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over time, the shorter pin may not engage in either the upper or lower body, causing the cord to come loose or slip … Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IL, KS, KY, LA, MA, ME, MN, MO, MS, NJ, NM, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA and the countries of Germany, Nethe…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
815 parts
Related Recalls
6
3 from same agency
Alpha Interlock Suspension System Lanyard Body: An external prosthetic device. The Alpha Lanyard includes suction with a one-way air expulsion valve plus a locking mechanism. Product design features an easy release mechanism. Available with or without a titanium pyramid. The lanyard is part of the Alpha Interlock" Suspension System. This system allows you to interchange suspension methods without remaking a socket Model, 3 part numbers - 700-AIS100, Alpha Interlock Suspension System with Alpha Lanyard - 700-AIS101, Alpha Interlock Suspension System with Alpha Lanyard w/Titanium Pyramid - 700-AIS107, Alpha Interlock Suspension System Lanyard Body ONLY
the firm discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over time, the shorter pin may not engage in either the upper or lower body, causing the cord to come loose or slip which could result in the loss of suspension.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1594-2021 |
| Date reported | May 26, 2021 |
| Date initiated | April 20, 2021 |
| Recalling firm | WillowWood Global Llc |
| Firm location | Mount Sterling, OH |
| Affected scope | 815 parts |
| Distribution | Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IL, KS, KY, LA, MA, ME, MN, MO, MS, NJ, NM, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA and the countries of Germany, Netherlands |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 26, 2021. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.