PlainRecalls
ModerateClass IIOngoing

FDA Devices recall · Reported September 14, 2022

VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461

Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction …

Recall #
Z-1676-2022
Affected scope
1148 units
Initiated
July 19, 2022
Verify with FDA Devices →
View my saved recalls

Ortho-Clinical Diagnostics, Inc. recalled VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461 - a moderate-severity action.

VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461 was recalled by Ortho-Clinical Diagnostics, Inc. in September 14, 2022. Reason: Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative sh…. Check the official notice for the remedy. Verify recall #Z-1676-2022 with the FDA Devices before acting.

The recall

Ortho-Clinical Diagnostics, Inc. issued this moderate-severity FDA Devices recall-Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative sh….

Moderate
severity level
1K units
affected scope
Class II
classification
September 14, 2022
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1676-2022 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461

Reason for recall

Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer. Potential exists for either a positive or negative shift in results reported by the VITROS System.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-1676-2022
Date reported September 14, 2022
Date initiated July 19, 2022
Recalling firm Ortho-Clinical Diagnostics, Inc.
Firm location Rochester, NY
Affected scope 1148 units
Distribution Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda DV04, Brazil, Canada L3R 4G5 , Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapo…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

1148 units units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1676-2022) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461. Recalled by Ortho-Clinical Diagnostics, Inc.. Units affected: 1148 units.
Why was this product recalled?
Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer. Potential exists for either a positive or negative shift in results reported by the VITROS System.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 14, 2022. Severity: Moderate. Recall number: Z-1676-2022.
Where was the recalled product distributed?
Distribution: Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda DV04, Brazil, Canada L3R 4G5 , Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported September 14, 2022. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026