Severity
Moderate
FDA Devices recall · Reported July 17, 2013
There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investigation of these reports indicat…
Baxter Healthcare Corp. recalled Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices th… - a moderate-severity action.
Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices th… was recalled by Baxter Healthcare Corp. in July 17, 2013. Reason: There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winge…. Check the official notice for the remedy. Verify recall #Z-1678-2013 with the FDA Devices before acting.
The recall
Baxter Healthcare Corp. issued this moderate-severity FDA Devices recall-There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winge….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1678-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1678-2013) was formally reported on July 17, 2013, with the manufacturer initiating the action on July 31, 2009. It is classified under Moderate severity (Class II), with a current status of Terminated. Baxter Healthcare Corp. is listed as the recalling firm, operating out of Deerfield, IL. Federal records list the affected scope as Code 2C1700KP: 7,128 units; Code 2C1702KP: 105,732 units.
The documented reason for this recall is: There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investigation of these reports indicates the incidence of leakage has only occ… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide), Puerto Rico and Internationally to Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, Columbia, Czech Republic, Denmark (including Faroe Islands), Ecuador, Finland, …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
Code 2C1700KP: 7,128 units; Code 2C1702KP: 105,732 units
Related Recalls
6
3 from same agency
Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices that deliver continuous medication for periods of 30 minutes up to 7 days. They are designed for medication therapies requiring continuous infusion. The infusion pumps can be utilized for slow, continuous delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), and subcutaneous or epidural infusion of medications directly into an intraoperative site or subcutaneously for post-operative pain management.
There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investigation of these reports indicates the incidence of leakage has only occurred after pumps are filled, primed, re-capped (using the blue winged Luer cap) and stored for a period of time, typically overnight. In addition, the reported instances of leakage
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1678-2013 |
| Date reported | July 17, 2013 |
| Date initiated | July 31, 2009 |
| Recalling firm | Baxter Healthcare Corp. |
| Firm location | Deerfield, IL |
| Affected scope | Code 2C1700KP: 7,128 units; Code 2C1702KP: 105,732 units |
| Distribution | Worldwide Distribution - USA (nationwide), Puerto Rico and Internationally to Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, Columbia, Czech Republic, Denmark (including Faroe Islands), Ecuador, Finland, France (including Ma… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported July 17, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.