Severity
Moderate
FDA Devices recall · Reported July 17, 2013
Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and C7122 APEX Arthroscopy Tubing Set, Two Connection, contained a scratc…
Linvatec Corp. dba ConMed Linvatec recalled ***REF C7122***QTY 1***APEX Arthroscopy Tubing Set. Intended for use for joint irrigat… - a moderate-severity action.
***REF C7122***QTY 1***APEX Arthroscopy Tubing Set. Intended for use for joint irrigat… was recalled by Linvatec Corp. dba ConMed Linvatec in July 17, 2013. Reason: Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and…. Check the official notice for the remedy. Verify recall #Z-1681-2013 with the FDA Devices before acting.
The recall
Linvatec Corp. dba ConMed Linvatec issued this moderate-severity FDA Devices recall-Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1681-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1681-2013) was formally reported on July 17, 2013, with the manufacturer initiating the action on May 30, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Linvatec Corp. dba ConMed Linvatec is listed as the recalling firm, operating out of Largo, FL. Federal records list the affected scope as 4,691 units, total of each.
The documented reason for this recall is: Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and C7122 APEX Arthroscopy Tubing Set, Two Connection, contained a scratch/grooved area on the inside diameter of… Distribution data in the federal record shows the product reached: Worldwide Distribution-USA (nationwide) including the states of AR, CA, FL, GA, IL, IN, KS, MO, NC, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, and WI, and the countries of Austria; Belgium, Canada, China, Czech Republic, F…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
4,691 units, total of each
Related Recalls
6
3 from same agency
***REF C7122***QTY 1***APEX Arthroscopy Tubing Set. Intended for use for joint irrigation and dissention during arthroscopic procedures.
Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and C7122 APEX Arthroscopy Tubing Set, Two Connection, contained a scratch/grooved area on the inside diameter of the pressure fitting that may cause the APEX Irrigation (pump) Console to sense a pressure drop and increase fluid flow to the surgical site.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1681-2013 |
| Date reported | July 17, 2013 |
| Date initiated | May 30, 2013 |
| Recalling firm | Linvatec Corp. dba ConMed Linvatec |
| Firm location | Largo, FL |
| Affected scope | 4,691 units, total of each |
| Distribution | Worldwide Distribution-USA (nationwide) including the states of AR, CA, FL, GA, IL, IN, KS, MO, NC, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, and WI, and the countries of Austria; Belgium, Canada, China, Czech Republic, Finland, France, Grea… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported July 17, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.