Severity
Moderate
FDA Devices recall · Reported April 12, 2017
Potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) a assay on the ADVIA Centaur Systems
Siemens Healthcare Diagnostics, Inc recalled Siemens ADVIA Centaur BR assay Cat No. 03896216, SMN 10334837 (5-pack/250 tests) BR… - a moderate-severity action.
Siemens ADVIA Centaur BR assay Cat No. 03896216, SMN 10334837 (5-pack/250 tests) BR… was recalled by Siemens Healthcare Diagnostics, Inc in April 12, 2017. Reason: Potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) a assay on the ADVIA Centa…. Check the official notice for the remedy. Verify recall #Z-1748-2017 with the FDA Devices before acting.
The recall
Siemens Healthcare Diagnostics, Inc issued this moderate-severity FDA Devices recall-Potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) a assay on the ADVIA Centa….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1748-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1748-2017) was formally reported on April 12, 2017, with the manufacturer initiating the action on March 8, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Siemens Healthcare Diagnostics, Inc is listed as the recalling firm, operating out of East Walpole, MA. Federal records list the affected scope as 1900 kits.
The documented reason for this recall is: Potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) a assay on the ADVIA Centaur Systems Distribution data in the federal record shows the product reached: Worldwide Distribution - US (Nationwide) Internationally to Australia, Brazil, Canada, Columbia, Cyprus, Japan, Spain. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
1900 kits
Related Recalls
6
3 from same agency
Siemens ADVIA Centaur BR assay Cat No. 03896216, SMN 10334837 (5-pack/250 tests) BR Assay for CA 27.29 The ADVIA Centaur¿ BR assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the ADVIA Centaur¿, ADVIA Centaur¿ XP, and ADVIA Centaur¿ XPT systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage III breast cancer. Serial testing for CA 27.29 in the serum of patients who are clinically free of disease should be used in conjunction with other clinical methods used for the early detection of cancer recurrence. The test is also intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring the progression or regression of disease in response to treatment.
Potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) a assay on the ADVIA Centaur Systems
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1748-2017 |
| Date reported | April 12, 2017 |
| Date initiated | March 8, 2017 |
| Recalling firm | Siemens Healthcare Diagnostics, Inc |
| Firm location | East Walpole, MA |
| Affected scope | 1900 kits |
| Distribution | Worldwide Distribution - US (Nationwide) Internationally to Australia, Brazil, Canada, Columbia, Cyprus, Japan, Spain |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 12, 2017. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.