Moderate
FDA Devices recall · Reported May 27, 2020
The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device may not turn on; 2) The device may not perform the selected function;…
Philips North America, LLC recalled Philips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for… - a moderate-severity action.
Philips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for… was recalled by Philips North America, LLC in May 27, 2020. Reason: The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device…. Check the official notice for the remedy. Verify recall #Z-2053-2020 with the FDA Devices before acting.
The recall
Philips North America, LLC issued this moderate-severity FDA Devices recall-The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2053-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2053-2020) was formally reported on May 27, 2020, with the manufacturer initiating the action on April 13, 2020. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips North America, LLC is listed as the recalling firm, operating out of Andover, MA. Federal records list the affected scope as 94,034 units.
The documented reason for this recall is: The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device may not turn on; 2) The device may not perform the selected function;3) The device may deliver a shock with a… Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
94,034 units
Related Recalls
6
3 from same agency
Philips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for use in the hospital by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support, or defibrillation. It must be used by or on the order of a physician.
The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device may not turn on; 2) The device may not perform the selected function;3) The device may deliver a shock with an energy level different than the setting selected by the user. In any of these scenarios, appropriate therapy delivery may be delayed.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2053-2020 |
| Date reported | May 27, 2020 |
| Date initiated | April 13, 2020 |
| Recalling firm | Philips North America, LLC |
| Firm location | Andover, MA |
| Affected scope | 94,034 units |
| Distribution | Worldwide distribution - US Nationwide distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 27, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.