FDA Devices recall · Reported August 29, 2012
Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Em…
Philips Medical Systems (Cleveland) Inc recalled Positron Emission Tomography (PET) and X-ray Computed Tomography, GEMINI TF 16 PET/CT, GE… - a moderate-severity action.
Positron Emission Tomography (PET) and X-ray Computed Tomography, GEMINI TF 16 PET/CT, GE… was recalled by Philips Medical Systems (Cleveland) Inc in August 29, 2012. Reason: Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole throug…. Check the official notice for the remedy. Verify recall #Z-2239-2012 with the FDA Devices before acting.
The recall
Philips Medical Systems (Cleveland) Inc issued this moderate-severity FDA Devices recall-Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole throug….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2239-2012 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2239-2012) was formally reported on August 29, 2012, with the manufacturer initiating the action on August 8, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips Medical Systems (Cleveland) Inc is listed as the recalling firm, operating out of Cleveland, OH. Federal records list the affected scope as 7 units.
The documented reason for this recall is: Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Emerson, determined the batteries, housed … Distribution data in the federal record shows the product reached: Nationwide Distribution-including the states of CO, FL, MD, MN, NE and PA.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
7 units
Related Recalls
6
3 from same agency
Positron Emission Tomography (PET) and X-ray Computed Tomography, GEMINI TF 16 PET/CT, GEMINI GXL 16 PET/CT, GEMINI TF Big Bore PET/CT, Philips Medical Systems, Cleveland, OH. The device is a diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data. The PET subsystem produces images of the distribution of PET radiopharmaceuticals in the patient body (specific radiopharmaceuticals are used for whole body, brain, heart and other organ imaging). Attenuation correction is accomplished by CTAC. The device also provides for list mode, dynamic, and gated acquisitions. Image processing and display workstations provide software applications to process, analyze, display, quantify and interpret medical images/data. The PET and CT images may be registered and displayed in a "fused" (overlaid in the same spatial orientation) format to provide combined metabolic and anatomical data at different angles. Trained professionals use the images in: " The evaluation, detection and diagnosis of lesions, disease and organ function such as but not limited to cancer, cardiovascular disease, and neurological disorders. " The detection, localization, and staging of tumors and diagnosing cancer patients. " Treatment planning and interventional radiology procedures. The device includes software that provides a quantified analysis of with regional cerebral activity from PET images.
Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Emerson, determined the batteries, housed within the UPS battery cabinet, entered thermal runaway causing the battery to overheat.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2239-2012 |
| Date reported | August 29, 2012 |
| Date initiated | August 8, 2012 |
| Recalling firm | Philips Medical Systems (Cleveland) Inc |
| Firm location | Cleveland, OH |
| Affected scope | 7 units |
| Distribution | Nationwide Distribution-including the states of CO, FL, MD, MN, NE and PA. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 29, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.