Severity
Moderate
Philips Medical Systems (Cleveland) Inc issued this FDA Devices recall on August 29, 2012. Classified as Moderate severity (Class II). Approximately 7 units units are affected. The recall was issued because: Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casi…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.
This FDA Devices action (record #Z-2239-2012) was formally reported on August 29, 2012, with the manufacturer initiating the action on August 8, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips Medical Systems (Cleveland) Inc is listed as the recalling firm, operating out of Cleveland, OH. Federal records indicate 7 units units are affected.
The documented reason for this recall is: Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Emerson, determined the batteries, housed … Distribution data in the federal record shows the product reached: Nationwide Distribution-including the states of CO, FL, MD, MN, NE and PA.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 14 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.
Severity
Moderate
Units Affected
7 units
Related Recalls
6
6 from same agency
Positron Emission Tomography (PET) and X-ray Computed Tomography, GEMINI TF 16 PET/CT, GEMINI GXL 16 PET/CT, GEMINI TF Big Bore PET/CT, Philips Medical Systems, Cleveland, OH. The device is a diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data. The PET subsystem produces images of the distribution of PET radiopharmaceuticals in the patient body (specific radiopharmaceuticals are used for whole body, brain, heart and other organ imaging). Attenuation correction is accomplished by CTAC. The device also provides for list mode, dynamic, and gated acquisitions. Image processing and display workstations provide software applications to process, analyze, display, quantify and interpret medical images/data. The PET and CT images may be registered and displayed in a "fused" (overlaid in the same spatial orientation) format to provide combined metabolic and anatomical data at different angles. Trained professionals use the images in: " The evaluation, detection and diagnosis of lesions, disease and organ function such as but not limited to cancer, cardiovascular disease, and neurological disorders. " The detection, localization, and staging of tumors and diagnosing cancer patients. " Treatment planning and interventional radiology procedures. The device includes software that provides a quantified analysis of with regional cerebral activity from PET images.
Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Emerson, determined the batteries, housed within the UPS battery cabinet, entered thermal runaway causing the battery to overheat.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2239-2012 |
| Date reported | August 29, 2012 |
| Date initiated | August 8, 2012 |
| Recalling firm | Philips Medical Systems (Cleveland) Inc |
| Units affected | 7 units |
| Distribution | Nationwide Distribution-including the states of CO, FL, MD, MN, NE and PA. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
Learn how the US recall system works and how to protect yourself and your household.
How the US Recall System Works
The three federal agencies, how recalls are initiated, and what happens next.
Understanding Recall Severity Classes
What Class I, II, and III mean and which recalls demand immediate action.
What to Do When a Product Is Recalled
Verify, claim your remedy, report injuries, and navigate the process.
How to Check If Your Products Are Recalled
Step-by-step guide to checking food, products, medications, and vehicles.
Recalled Products in Your Home
A room-by-room household audit guide for active recalls.
Most Recalled Product Categories
Rankings of highest-recall categories from FDA, CPSC, and NHTSA.
Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.
Explore related product safety and public health data from federal sources.
Food Safety Inspections
FDA and state food facility inspection results — violations, enforcement actions, and compliance history for food manufacturers and processors.
PlainFoodSafe →
Ingredient Safety Data
FDA food additive and ingredient safety database — regulatory status, usage limits, and safety assessments for thousands of ingredients.
PlainIngredients →
Nutrition Data
USDA nutrition facts for 2M+ food items — calories, macros, vitamins, and ingredient analysis to verify what you consume.
GetFoodFacts →
Drug Safety Information
FDA drug data for 680+ medications — interactions, alternatives, side effects, and safety information for recalled and active drugs.
PlainMeds →
Product Injury Data
CPSC emergency room injury data for 838 product categories — 7.3M NEISS records tracking real-world consumer product injuries.
PlainSafety →
Recall Checker
Search our database of 83K+ recalls by product name, brand, or recall number across all agencies.
Check Now →
Recall Radar
Live feed of the latest recalls across FDA, CPSC, NHTSA, and USDA — filter by agency and severity.
View Radar →
Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.
Waldemar Link GmbH & Co. KG (Mfg Site) · 2026-03-18
Penner Patient Care, Inc. · 2026-03-18
Instrumentation Laboratory · 2026-03-18
Beckman Coulter Mishima K.K. · 2026-03-18
Trividia Health, Inc. · 2026-03-18
Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →
Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.
Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.
Read our methodology — how this data is sourced, computed, and verified.
Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).