FDA Devices recall · Reported August 29, 2012
GE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analysis showing that a replacement Hitachi 160 GB Hard Drive, when install…
Ge Oec Medical Systems, Inc recalled 892.1650 Image-intensified fluoroscopic x-ray system; the device is designed to provide f… - a moderate-severity action.
892.1650 Image-intensified fluoroscopic x-ray system; the device is designed to provide f… was recalled by Ge Oec Medical Systems, Inc in August 29, 2012. Reason: GE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analy…. Check the official notice for the remedy. Verify recall #Z-2240-2012 with the FDA Devices before acting.
The recall
Ge Oec Medical Systems, Inc issued this moderate-severity FDA Devices recall-GE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analy….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2240-2012 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2240-2012) was formally reported on August 29, 2012, with the manufacturer initiating the action on November 6, 2009. It is classified under Moderate severity (Class II), with a current status of Terminated. Ge Oec Medical Systems, Inc is listed as the recalling firm, operating out of Salt Lake City, UT. Federal records list the affected scope as 3,595 units.
The documented reason for this recall is: GE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analysis showing that a replacement Hitachi 160 GB Hard Drive, when installed on certain models is susceptible of c… Distribution data in the federal record shows the product reached: Worldwide Distribution -- USA (nationwide). Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
3,595 units
Related Recalls
6
3 from same agency
892.1650 Image-intensified fluoroscopic x-ray system; the device is designed to provide fluoroscopic and spot film imaging of the patient during diagnostic surgical and interventional procedures. The systems include features specifically designed for use in diagnostic and interventional cardiac imaging procedures, and are also intended for cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, critical care and emergency room procedures. The systems may be used for other imaging applications at the physician's discretion.
GE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analysis showing that a replacement Hitachi 160 GB Hard Drive, when installed on certain models is susceptible of causing unanticipated system shut downs, no boots, data loss or data mix, or unexpected hard drive corruption.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2240-2012 |
| Date reported | August 29, 2012 |
| Date initiated | November 6, 2009 |
| Recalling firm | Ge Oec Medical Systems, Inc |
| Firm location | Salt Lake City, UT |
| Affected scope | 3,595 units |
| Distribution | Worldwide Distribution -- USA (nationwide) |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 29, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.