PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported July 27, 2016

cobas EGFR Mutation Test V2 Materials number CE-IVD: EGFR v2: 07248563190 cfDNA: 07247737190 Device Identifier: EGFR v2.0: 00875197005448 cfDNA: 00875197005424 Product Usage: A real time PCR Test for the qualitative detection and identifications of mutations n exons 18, 19, 20 and 21 of the epidermal growth factor receptor (EGFR) gene in DNA derived from formalin-fixed paraffin-embedded (FFPET) tumor tissue or plasma from non-small cell lung cancer (NSCLC) patients. The test also is

During internal studies using contrived plasma specimens, several mutations (L8568R, Exon 19 deletion, T790M) inconsistently generated "No Mutation Detected" (ie False Negative ) …

Recall #
Z-2254-2016
Affected scope
735 devices EGFR; 509 devices cfDNA
Initiated
April 5, 2016
Verify with FDA Devices →
View my saved recalls

Roche Molecular Systems, Inc. recalled cobas EGFR Mutation Test V2 Materials number CE-IVD: EGFR v2: 07248563190 cfDNA: 0724… - a moderate-severity action.

cobas EGFR Mutation Test V2 Materials number CE-IVD: EGFR v2: 07248563190 cfDNA: 0724… was recalled by Roche Molecular Systems, Inc. in July 27, 2016. Reason: During internal studies using contrived plasma specimens, several mutations (L8568R, Exon 19 deletion, T790M)…. Check the official notice for the remedy. Verify recall #Z-2254-2016 with the FDA Devices before acting.

The recall

Roche Molecular Systems, Inc. issued this moderate-severity FDA Devices recall-During internal studies using contrived plasma specimens, several mutations (L8568R, Exon 19 deletion, T790M)….

Moderate
severity level
735 units
affected scope
Class II
classification
July 27, 2016
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-2254-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

cobas EGFR Mutation Test V2 Materials number CE-IVD: EGFR v2: 07248563190 cfDNA: 07247737190 Device Identifier: EGFR v2.0: 00875197005448 cfDNA: 00875197005424 Product Usage: A real time PCR Test for the qualitative detection and identifications of mutations n exons 18, 19, 20 and 21 of the epidermal growth factor receptor (EGFR) gene in DNA derived from formalin-fixed paraffin-embedded (FFPET) tumor tissue or plasma from non-small cell lung cancer (NSCLC) patients. The test also is intended to be used as an aid in selecting patients with NSCLC for therapy with an EGFR tyrosine kinase inhibitor (TKI).

Reason for recall

During internal studies using contrived plasma specimens, several mutations (L8568R, Exon 19 deletion, T790M) inconsistently generated "No Mutation Detected" (ie False Negative ) result with the cobas EGFR mutation Text V2.0 when utilizing the cobas cfDNA Sample Preparation kit.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-2254-2016
Date reported July 27, 2016
Date initiated April 5, 2016
Recalling firm Roche Molecular Systems, Inc.
Firm location Branchburg, NJ
Affected scope 735 devices EGFR; 509 devices cfDNA
Distribution Foriegn Distribution Austria, Belgium, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hungary , Italy, Lithuania, Luxembourg, Netherlands, Norway, Poland, Romania, Slovakia, Slovenia, Spain, Switzerland, Turkey, Unit…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

735 devices EGFR; 509 devices cfDNA units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-2254-2016) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
cobas EGFR Mutation Test V2 Materials number CE-IVD: EGFR v2: 07248563190 cfDNA: 07247737190 Device Identifier: EGFR v2.0: 00875197005448 cfDNA: 00875197005424 Product Usage: A real time PCR Test for the qualitative detection and identifications of mutations n exons 18, 19, 20 and 21 of the epidermal growth factor receptor (EGFR) gene in DNA derived from formalin-fixed paraffin-embedded (FFPET) tumor tissue or plasma from non-small cell lung cancer (NSCLC) patients. The test also is intended to be used as an aid in selecting patients with NSCLC for therapy with an EGFR tyrosine kinase inhibitor (TKI).. Recalled by Roche Molecular Systems, Inc.. Units affected: 735 devices EGFR; 509 devices cfDNA.
Why was this product recalled?
During internal studies using contrived plasma specimens, several mutations (L8568R, Exon 19 deletion, T790M) inconsistently generated "No Mutation Detected" (ie False Negative ) result with the cobas EGFR mutation Text V2.0 when utilizing the cobas cfDNA Sample Preparation kit.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 27, 2016. Severity: Moderate. Recall number: Z-2254-2016.
Where was the recalled product distributed?
Distribution: Foriegn Distribution Austria, Belgium, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hungary , Italy, Lithuania, Luxembourg, Netherlands, Norway, Poland, Romania, Slovakia, Slovenia, Spain, Switzerland, Turkey, United Kingdom cfDNA: Austria, Belgium, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hungary , Italy, Lithuania, Luxembourg, Netherlands, Poland, Romania, Slovakia, Slovenia, Spain, Switzerland, Turkey, United Kingdom.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported July 27, 2016. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026