PlainRecalls
FDA Devices Moderate Class II Terminated

AUTOCOMP6 High Speed Compounder

Reported: August 3, 2016 Initiated: June 23, 2016 #Z-2259-2016

Product Description

AUTOCOMP6 High Speed Compounder

Reason for Recall

An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.

Details

Recalling Firm
The Metrix Company
Units Affected
323
Distribution
Product distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United Kingdom
Location
Dubuque, IA

Frequently Asked Questions

What product was recalled?
AUTOCOMP6 High Speed Compounder. Recalled by The Metrix Company. Units affected: 323.
Why was this product recalled?
An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 3, 2016. Severity: Moderate. Recall number: Z-2259-2016.