FDA Devices recall · Reported August 3, 2016
Imagecast PACS with Centricity RIS-IC versions prior to 10.6 Update Package 18 Product Usage: is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging devices and modalities. Lossy compressed mam
A software defect was discovered that causes images to be out of context with clinical information.
Ge Healthcare It recalled Imagecast PACS with Centricity RIS-IC versions prior to 10.6 Update Package 18 Product U… - a moderate-severity action.
Imagecast PACS with Centricity RIS-IC versions prior to 10.6 Update Package 18 Product U… was recalled by Ge Healthcare It in August 3, 2016. Reason: A software defect was discovered that causes images to be out of context with clinical information.. Check the official notice for the remedy. Verify recall #Z-2300-2016 with the FDA Devices before acting.
The recall
Ge Healthcare It issued this moderate-severity FDA Devices recall-A software defect was discovered that causes images to be out of context with clinical information..
- Moderate
- severity level
- 13 units
- affected scope
- Class II
- classification
- August 3, 2016
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-2300-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Imagecast PACS with Centricity RIS-IC versions prior to 10.6 Update Package 18 Product Usage: is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging devices and modalities. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by the FDA.
Reason for recall
A software defect was discovered that causes images to be out of context with clinical information.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2300-2016 |
| Date reported | August 3, 2016 |
| Date initiated | February 18, 2013 |
| Recalling firm | Ge Healthcare It |
| Firm location | Barrington, IL |
| Affected scope | 13 |
| Distribution | US Nationwide Distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-2300-2016) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026