Alere INRatio PT/INR Test Strips Model Number: 0100071, 0100139 Product Usage: The INRatio/INRatio2 Test Strips perform a modified version of the one-stage Prothrombin Time test, using a recombinant human thromboplastin reagent. The clot formed in the Prothrombin Time reaction is detected by a change in the electrical impedance of the sample during the coagulation process. The system consists of a monitor and disposable test strips. The monitor measures impedance, heats the test strip to th

Reported: August 31, 2016 Initiated: July 11, 2016 #Z-2361-2016

Product Description

Alere INRatio PT/INR Test Strips Model Number: 0100071, 0100139 Product Usage: The INRatio/INRatio2 Test Strips perform a modified version of the one-stage Prothrombin Time test, using a recombinant human thromboplastin reagent. The clot formed in the Prothrombin Time reaction is detected by a change in the electrical impedance of the sample during the coagulation process. The system consists of a monitor and disposable test strips. The monitor measures impedance, heats the test strip to the proper reaction temperature, and provides a user interface. The blood sample is applied to the Test Strip and the clotting reaction occurs on the Test Strip.

Reason for Recall

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Details

Recalling Firm
Alere San Diego, Inc.
Distribution
Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Malaysia, Maldives, Netherlands, Russia, Rwanda, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad, Ukraine, Uruguay, and Vietnam.
Location
San Diego, CA

Frequently Asked Questions

What product was recalled?
Alere INRatio PT/INR Test Strips Model Number: 0100071, 0100139 Product Usage: The INRatio/INRatio2 Test Strips perform a modified version of the one-stage Prothrombin Time test, using a recombinant human thromboplastin reagent. The clot formed in the Prothrombin Time reaction is detected by a change in the electrical impedance of the sample during the coagulation process. The system consists of a monitor and disposable test strips. The monitor measures impedance, heats the test strip to the proper reaction temperature, and provides a user interface. The blood sample is applied to the Test Strip and the clotting reaction occurs on the Test Strip.. Recalled by Alere San Diego, Inc..
Why was this product recalled?
Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 31, 2016. Severity: Critical. Recall number: Z-2361-2016.