CriticalClass ITerminated

FDA Devices recall · Reported August 31, 2016

Alere INRatio/ INRatio2 PT/INR Test Strips Model Number: 99007EU, 99007G1, 99007G3, 99007G5, 99007G7, 99008EU, 99008G1, 99008G3, 99008G5, 99008G7 Product Usage: The Alere INRatio¿/INRatio¿2 system uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that initiate coagulation. When the blood clots, changes in impedance in the sample are detected by the monitor. The monitor calculates the P

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Test Strips. This action is being initiated as a result of patients who had a therapeutic …

Recall #
Z-2362-2016
Initiated
July 11, 2016
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Alere San Diego, Inc. recalled Alere INRatio/ INRatio2 PT/INR Test Strips Model Number: 99007EU, 99007G1, 99007G3, 9900… - a critical-severity action.

Alere INRatio/ INRatio2 PT/INR Test Strips Model Number: 99007EU, 99007G1, 99007G3, 9900… was recalled by Alere San Diego, Inc. in August 31, 2016. Reason: Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Test Strips. This ac…. Check the official notice for the remedy. Verify recall #Z-2362-2016 with the FDA Devices before acting.

The recall

Alere San Diego, Inc. issued this critical-severity FDA Devices recall-Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Test Strips. This ac….

Critical
severity level
Class I
classification
August 31, 2016
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-2362-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-2362-2016) was formally reported on August 31, 2016, with the manufacturer initiating the action on July 11, 2016. It is classified under Critical severity (Class I), with a current status of Terminated. Alere San Diego, Inc. is listed as the recalling firm, operating out of San Diego, CA. The number of affected units is not disclosed in the agency filing, which is common for drug and food recalls where lot-level tracking supersedes unit counts.

The documented reason for this recall is: Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Test Strips. This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere … Distribution data in the federal record shows the product reached: Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified high severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Critical

Affected scope

Unknown

Related Recalls

6

3 from same agency

Product description

Alere INRatio/ INRatio2 PT/INR Test Strips Model Number: 99007EU, 99007G1, 99007G3, 99007G5, 99007G7, 99008EU, 99008G1, 99008G3, 99008G5, 99008G7 Product Usage: The Alere INRatio¿/INRatio¿2 system uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that initiate coagulation. When the blood clots, changes in impedance in the sample are detected by the monitor. The monitor calculates the PT and INR results from this impedance change and reports them on the display. The Alere INRatio/INRatio2 PT/INR Monitoring System is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for use outside the body (in vitro diagnostic use). The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for professional and home use by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is not intended to be used for screening purposes.

Reason for recall

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Test Strips. This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Critical (Class I)
Status Terminated
Recall number Z-2362-2016
Date reported August 31, 2016
Date initiated July 11, 2016
Recalling firm Alere San Diego, Inc.
Firm location San Diego, CA
Affected scope Not disclosed
Distribution Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, In…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-2362-2016) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • This is a Class I (high-risk) recall, stop using the product immediately and follow the disposal or return instructions. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Alere INRatio/ INRatio2 PT/INR Test Strips Model Number: 99007EU, 99007G1, 99007G3, 99007G5, 99007G7, 99008EU, 99008G1, 99008G3, 99008G5, 99008G7 Product Usage: The Alere INRatio¿/INRatio¿2 system uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that initiate coagulation. When the blood clots, changes in impedance in the sample are detected by the monitor. The monitor calculates the PT and INR results from this impedance change and reports them on the display. The Alere INRatio/INRatio2 PT/INR Monitoring System is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for use outside the body (in vitro diagnostic use). The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for professional and home use by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is not intended to be used for screening purposes.. Recalled by Alere San Diego, Inc..
Why was this product recalled?
Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Test Strips. This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 31, 2016. Severity: Critical. Recall number: Z-2362-2016.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Malaysia, Maldives, Netherlands, Russia, Rwanda, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad, Ukraine, Uruguay, and Vietnam..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-2362-2016) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 31, 2016. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.