Severity
Moderate
FDA Devices recall · Reported September 3, 2025
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage app…
Dentsply Ih, Inc. recalled Brand Name: Atlantis IO FLO-S Product Name: Atlantis IO FLO-S Kit Model/Catalog Number:… - a moderate-severity action.
Brand Name: Atlantis IO FLO-S Product Name: Atlantis IO FLO-S Kit Model/Catalog Number:… was recalled by Dentsply Ih, Inc. in September 3, 2025. Reason: The defect in the thread area will not allow the device to fully engage with an implant or analog (implant re…. Check the official notice for the remedy. Verify recall #Z-2421-2025 with the FDA Devices before acting.
The recall
Dentsply Ih, Inc. issued this moderate-severity FDA Devices recall-The defect in the thread area will not allow the device to fully engage with an implant or analog (implant re….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2421-2025 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2421-2025) was formally reported on September 3, 2025, with the manufacturer initiating the action on July 24, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Dentsply Ih, Inc. is listed as the recalling firm, operating out of Waltham, MA. Federal records list the affected scope as 102 units.
The documented reason for this recall is: The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to ful… Distribution data in the federal record shows the product reached: Worldwide - US Nationwide distribution in the states of IL, LA, MA, MN, MO, ND, NJ, NM, OH, OK, PA, TX, WV and the countries of AT, AU, BE, CA, DE, DK, ES, FI, FR, IT, JP, NL, NO, PL, QA, SE, UA, UK.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
102 units
Related Recalls
6
3 from same agency
Brand Name: Atlantis IO FLO-S Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04B / 68020033 Software Version: n/a Product Description: The Atlantis intraoral feature locating object for suprastructures (IO FLO-S) is intended to support the prosthetic procedure with the aim of restoring chewing function in partial or complete edentulous patients, by enabling digital impression taking on implant or abutment level for all positions in the mouth of the patient or on the master model. Component: not a component
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-2421-2025 |
| Date reported | September 3, 2025 |
| Date initiated | July 24, 2025 |
| Recalling firm | Dentsply Ih, Inc. |
| Firm location | Waltham, MA |
| Affected scope | 102 units |
| Distribution | Worldwide - US Nationwide distribution in the states of IL, LA, MA, MN, MO, ND, NJ, NM, OH, OK, PA, TX, WV and the countries of AT, AU, BE, CA, DE, DK, ES, FI, FR, IT, JP, NL, NO, PL, QA, SE, UA, UK. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 3, 2025. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.