Severity
Moderate
FDA Devices recall · Reported August 14, 2024
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the fol…
Ge Medical Systems, LLC recalled SIGNA Victor, with affected software versions: MR30.1; Nuclear Magnetic Resonance Imagin… - a moderate-severity action.
SIGNA Victor, with affected software versions: MR30.1; Nuclear Magnetic Resonance Imagin… was recalled by Ge Medical Systems, LLC in August 14, 2024. Reason: GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the …. Check the official notice for the remedy. Verify recall #Z-2480-2024 with the FDA Devices before acting.
The recall
Ge Medical Systems, LLC issued this moderate-severity FDA Devices recall-GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2480-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2480-2024) was formally reported on August 14, 2024, with the manufacturer initiating the action on June 14, 2024. It is classified under Moderate severity (Class II), with a current status of Ongoing. Ge Medical Systems, LLC is listed as the recalling firm, operating out of Waukesha, WI. Federal records list the affected scope as 23 units.
The documented reason for this recall is: GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FS… Distribution data in the federal record shows the product reached: Worldwide distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
23 units
Related Recalls
6
3 from same agency
SIGNA Victor, with affected software versions: MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-2480-2024 |
| Date reported | August 14, 2024 |
| Date initiated | June 14, 2024 |
| Recalling firm | Ge Medical Systems, LLC |
| Firm location | Waukesha, WI |
| Affected scope | 23 units |
| Distribution | Worldwide distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 14, 2024. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.