PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported September 2, 2020

Cios Alpha (VA20) - Model: 10308191 Cios Alpha (VA30) - Model: 11105200 Cios Spin (VA30) - Model: 10308194 Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications

Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (…

Recall #
Z-2848-2020
Affected scope
US: 282 units; OUS: 1363
Initiated
July 14, 2020
Verify with FDA Devices →
View my saved recalls

Siemens Medical Solutions USA, Inc recalled Cios Alpha (VA20) - Model: 10308191 Cios Alpha (VA30) - Model: 11105200 Cios Spin (VA30… - a moderate-severity action.

Cios Alpha (VA20) - Model: 10308191 Cios Alpha (VA30) - Model: 11105200 Cios Spin (VA30… was recalled by Siemens Medical Solutions USA, Inc in September 2, 2020. Reason: Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below re…. Check the official notice for the remedy. Verify recall #Z-2848-2020 with the FDA Devices before acting.

The recall

Siemens Medical Solutions USA, Inc issued this moderate-severity FDA Devices recall-Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below re….

Moderate
severity level
Class II
classification
September 2, 2020
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-2848-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Cios Alpha (VA20) - Model: 10308191 Cios Alpha (VA30) - Model: 11105200 Cios Spin (VA30) - Model: 10308194 Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications

Reason for recall

Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of imaging functionality.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-2848-2020
Date reported September 2, 2020
Date initiated July 14, 2020
Recalling firm Siemens Medical Solutions USA, Inc
Firm location Malvern, PA
Affected scope US: 282 units; OUS: 1363
Distribution Nationwide

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-2848-2020) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Cios Alpha (VA20) - Model: 10308191 Cios Alpha (VA30) - Model: 11105200 Cios Spin (VA30) - Model: 10308194 Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Recalled by Siemens Medical Solutions USA, Inc. Units affected: US: 282 units; OUS: 1363.
Why was this product recalled?
Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of imaging functionality.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 2, 2020. Severity: Moderate. Recall number: Z-2848-2020.
Where was the recalled product distributed?
Distribution: Nationwide.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported September 2, 2020. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026