Cios Alpha (VA20) - Model: 10308191 Cios Alpha (VA30) - Model: 11105200 Cios Spin (VA30) - Model: 10308194 Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications
Reported: September 2, 2020 Initiated: July 14, 2020 #Z-2848-2020
Product Description
Cios Alpha (VA20) - Model: 10308191 Cios Alpha (VA30) - Model: 11105200 Cios Spin (VA30) - Model: 10308194 Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications
Reason for Recall
Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of imaging functionality.
Details
- Recalling Firm
- Siemens Medical Solutions USA, Inc
- Units Affected
- US: 282 units; OUS: 1363
- Distribution
- Nationwide
- Location
- Malvern, PA
Frequently Asked Questions
What product was recalled? ▼
Cios Alpha (VA20) - Model: 10308191 Cios Alpha (VA30) - Model: 11105200 Cios Spin (VA30) - Model: 10308194 Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Recalled by Siemens Medical Solutions USA, Inc. Units affected: US: 282 units; OUS: 1363.
Why was this product recalled? ▼
Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of imaging functionality.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on September 2, 2020. Severity: Moderate. Recall number: Z-2848-2020.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11