PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported September 30, 2020

Inquiry Steerable Diagnostic Catheter, REF: IBI-81104 - Product Usage: INDICATIONS FOR USE The Inquiry fixed curve and steerable electrophysiology catheters are used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies. The catheters are commonly placed at the high right atrium, right ventricular apex, and HIS bundle.

Steerable diagnostic catheter packages labeled Model IBI-81104, batch 7397352 incorrectly contained Model IBI-87002, batch 7397307. Packages labeled Model IBI-87002, batch 7397307…

Recall #
Z-3011-2020
Affected scope
15 units
Initiated
August 24, 2020
Verify with FDA Devices →
View my saved recalls

Irvine Biomedical Inc, a St. Jude Medical Co. recalled Inquiry Steerable Diagnostic Catheter, REF: IBI-81104 - Product Usage: INDICATIONS FOR US… - a moderate-severity action.

Inquiry Steerable Diagnostic Catheter, REF: IBI-81104 - Product Usage: INDICATIONS FOR US… was recalled by Irvine Biomedical Inc, a St. Jude Medical Co. in September 30, 2020. Reason: Steerable diagnostic catheter packages labeled Model IBI-81104, batch 7397352 incorrectly contained Model IBI…. Check the official notice for the remedy. Verify recall #Z-3011-2020 with the FDA Devices before acting.

The recall

Irvine Biomedical Inc, a St. Jude Medical Co. issued this moderate-severity FDA Devices recall-Steerable diagnostic catheter packages labeled Model IBI-81104, batch 7397352 incorrectly contained Model IBI….

Moderate
severity level
15 units
affected scope
Class II
classification
September 30, 2020
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-3011-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Inquiry Steerable Diagnostic Catheter, REF: IBI-81104 - Product Usage: INDICATIONS FOR USE The Inquiry fixed curve and steerable electrophysiology catheters are used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies. The catheters are commonly placed at the high right atrium, right ventricular apex, and HIS bundle.

Reason for recall

Steerable diagnostic catheter packages labeled Model IBI-81104, batch 7397352 incorrectly contained Model IBI-87002, batch 7397307. Packages labeled Model IBI-87002, batch 7397307 incorrectly contained Model IBI-81104, batch 7397352.If physicians fail to observe the catheter difference, the intracardiac electrograms will not be representative of the desired signals, could result in procedure delay

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-3011-2020
Date reported September 30, 2020
Date initiated August 24, 2020
Recalling firm Irvine Biomedical Inc, a St. Jude Medical Co.
Firm location Irvine, CA
Affected scope 15 units
Distribution Worldwide distribution - U.S. Nationwide distribution including the states of MO, IN, WI, TX, AZ, MI, VA, KY and the country of Japan.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

15 units units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-3011-2020) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Inquiry Steerable Diagnostic Catheter, REF: IBI-81104 - Product Usage: INDICATIONS FOR USE The Inquiry fixed curve and steerable electrophysiology catheters are used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies. The catheters are commonly placed at the high right atrium, right ventricular apex, and HIS bundle.. Recalled by Irvine Biomedical Inc, a St. Jude Medical Co.. Units affected: 15 units.
Why was this product recalled?
Steerable diagnostic catheter packages labeled Model IBI-81104, batch 7397352 incorrectly contained Model IBI-87002, batch 7397307. Packages labeled Model IBI-87002, batch 7397307 incorrectly contained Model IBI-81104, batch 7397352.If physicians fail to observe the catheter difference, the intracardiac electrograms will not be representative of the desired signals, could result in procedure delay
Which agency issued this recall?
This recall was issued by the FDA Devices on September 30, 2020. Severity: Moderate. Recall number: Z-3011-2020.
Where was the recalled product distributed?
Distribution: Worldwide distribution - U.S. Nationwide distribution including the states of MO, IN, WI, TX, AZ, MI, VA, KY and the country of Japan..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported September 30, 2020. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026