Severity
Moderate
FDA Devices recall · Reported November 21, 2012
Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using th…
Terumo Cardiovascular Systems Corporation recalled HX2 Temp Manage SYS 120V The Terumo HX2 Temperature Management System is a selfcontain… - a moderate-severity action.
HX2 Temp Manage SYS 120V The Terumo HX2 Temperature Management System is a selfcontain… was recalled by Terumo Cardiovascular Systems Corporation in November 21, 2012. Reason: Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temper…. Check the official notice for the remedy. Verify recall #Z-0278-2013 with the FDA Devices before acting.
The recall
Terumo Cardiovascular Systems Corporation issued this moderate-severity FDA Devices recall-Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temper….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0278-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0278-2013) was formally reported on November 21, 2012, with the manufacturer initiating the action on October 16, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Terumo Cardiovascular Systems Corporation is listed as the recalling firm, operating out of Ann Arbor, MI. Federal records list the affected scope as 19 units.
The documented reason for this recall is: Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the i… Distribution data in the federal record shows the product reached: Nationwide Distribution including CA, HI, KS, LA, MI, MO, OR, PA, UT, and WI. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
19 units
Related Recalls
6
3 from same agency
HX2 Temp Manage SYS 120V The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit.
Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0278-2013 |
| Date reported | November 21, 2012 |
| Date initiated | October 16, 2012 |
| Recalling firm | Terumo Cardiovascular Systems Corporation |
| Firm location | Ann Arbor, MI |
| Affected scope | 19 units |
| Distribution | Nationwide Distribution including CA, HI, KS, LA, MI, MO, OR, PA, UT, and WI |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 21, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.