FDA Devices recall · Reported November 21, 2012
PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures.
Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical…
Dako North America Inc. recalled PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standa… - a moderate-severity action.
PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standa… was recalled by Dako North America Inc. in November 21, 2012. Reason: Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heate…. Check the official notice for the remedy. Verify recall #Z-0287-2013 with the FDA Devices before acting.
The recall
Dako North America Inc. issued this moderate-severity FDA Devices recall-Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heate….
- Moderate
- severity level
- 136 units
- affected scope
- Class II
- classification
- November 21, 2012
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-0287-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures.
Reason for recall
Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0287-2013 |
| Date reported | November 21, 2012 |
| Date initiated | May 1, 2008 |
| Recalling firm | Dako North America Inc. |
| Firm location | Carpinteria, CA |
| Affected scope | 136 units |
| Distribution | Nationwide distribution: USA including states of: AK, CA, CO, CT, DC, FL, GA, HI, IA, IL IN, KS, KY, MA, MD, MI, MN, MO, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, and WA. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-0287-2013) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls